Feature
- Accepts syringe from 2/3 to 60 ml
- Customized drug library (up to 300 drug labels)
- Highly visible pump status indicator
- Easy intuitive programming and operation
- Automatic capturing of syringe plungers
- Improved alarms signals
- Graphic LCD dispaly
- Rate range from 0.01 to 2200 ml/h
- Advanced cummunications ( USB) , Irda Interfaces )
Performance
Flow Rate | : | 0.01 – 99,99 ml/h in 0.01 ml/h steps |
100.0 â€999.9 ml/h in 0.1 ml/h steps | ||
1000 †2200 ml/h in 1 ml/h steps | ||
Mechanical accuracy | : | ± 1% |
Volumetric accuracy with | : | ± 2% or ± 0.1 ml/h (whichever is greater) |
approved syringes | ||
Compatible syringes, ml | : | 2/3, 5, 10, 20, 30, and 50/60 of all major brands |
Occulusion pressure ( adjustable ) | : | L 1 ( approx. 50 mmHg ) - L10 (approx. 950 mmHg) |
Dose Units | : | g, mg, mcg, ng, Units, kUnits, mUnits, mcUnits, mol , |
mmol , mcmol , mEq, and other configured by the medical personel | ||
Purge volume | : | 0.1 – 4.0 ml in steps of 0.1 ml |
Bolus/purge rates | : | 1 – 2200 ml /h in steps of 1 ml/h |
Bolus Volume | : | 0.1 – 60 ml in 0.1 ml step |
Keep Vein Open ( KVO ) rate | : | 0.1 ml/h – 5 ml/h |
Purge rates | : | 1 – 2200 ml /h in steps of 1 ml/h |
KVO Volume | : | 0.1 % †10% of the syringe Volume |
Volume To be Infused ( VTBI ) | : | 0.1 - 99.9 ml in steps of 0.01 ml, 100 - 999 ml in steps of |
0.1 ml, 1000 - 9999 ml in steps of 1 ml or No Limit | ||
Post VTBI rate | : | Stop , KVO or continue with set rate |
Standby time | : | 1 min – 24 h |
Alarms | : | Visual and audio |
Dose error reduction | : | Infugard protocol ( with safety limits ) |
Interfaces | : | USB ( peripheral ), IrDA ( IR comms port ) |
Mounting | : | Table top operation |
Universal pole clamps | ||
Draeger (horizontal) bar | ||
Infusion device rack | ||
Power supply | : | 100 – 240VAC , 50/60 Hz, 50 VA or |
internal rechargeable battery | ||
Battery type | : | Built-in NiMH 7.2V/2.5 Ah |
Battery life | : | 10 h @ 5 ml/h |
Battery charging time | : | 5 hours to 100% charge |
Event Log | : | > 2000 events |
History log | : | > 500 events |
Key log | : | > last 300 presses |