Product >> Instramed >> ISIS PRO
ISIS PRO

AUTOMATED EXTERNAL DEFIBRILLATOR WITH MANUAL FUNCTION 

ISIS PRO

 

FEATURE

  • Double use: as manual defirillator or AED
  • Audio orientation and visual indicators on the LCD screen
  • Horizontal or upright mount
  • Internal memory torecord events
  • USB-PC connection, includes software for data download and management
  • Self-diagnosis of functions and battery
  • Easy access to pads, for use and replacement

 

GENERAL SPECIFICATIONS

  • Dimensions

- 22.0 cm / 8.66 in (W)
- 13.0 cm / 5.11 in (D)
- 29.0 cm / 11.41 in (H)

  • Weight

- Device 2.90 Kg / 6.39 lb

  • Internal rechargeable battery

- Type: Li-ion, 14.4 VDC 4.0 A/h

  • Duration

- 10 hours in cardiac rhythm recognition mode (full load battery) or at least 200 shocks
in 200 Joules (battery under good conditions)
- Battery full load time (fully unloaded): 5 hours

  • Battery loader source

- Power supply 100 - 240V/50 -60 Hz
- Consumption (max): Power supply 1 A.
- Output: 24 VDC, 1.5A

  • Battery storage:

- Battery storage for long periods under temperatures above 35oC will reduce the
battery's capacity and abbreviative its useful life

  • Preset defibrillation scales (AED mode):

- Adult: 1st shock 150 J, subsequent shocks 200 J
- Child: 50 J

  • Internal memory storage:

- 100 events or 2 hours of ECG record

  • Protection class:

- IPXO

  • Classification:

- Internal Energized Equipment
- CF Type

  • Mode of operation

- Continuous Operation

  • Maximum time since beginning of rhythm analysis until promptness for discharged:

- 20 s

  • Maximum time since beginning of the operation untul promptness for discharge at the maximum energy:

- 25 s

 

ENVIRONMENTAL SPECIFICATIONS

  • Temperature:

- Operational: 0 to 50oC/32 to 122 F
- Storage: 0 to 70oC/32 to 158 F

  • Humidity:

- Operational: 10 to 95% RH, without condensation
- Storage: 10 to 95% RH, without condensation

 

DEFIBRILLATOR

  • Wave form:

- Biphasic truncated Exponential. Wave form parameters set according to the patient's
impedance

  • Shock application:

- By Multifunctional adhesive pads

  • Commands:

- Front Panel button (on/off)
- Touchscreen: allows for the selection of manual made and defining energy scales

  • Defibrillation scales:

- Adult: 150 to 270 J
- Child: 10, 20, 30, 40 and 50 J

  • Adult / Child Selection:

- Automatic by the type of pads

  • Charge command:

- Automatic after identifying arrhytmias that should receive shock

  • Shock command

- Front Panel button, when blinking

  • Maximum time between beginning of rhythm analysis and promptness for discharge

- 20 s

  • Maximum charging time

- 50 Joules: < 2 seconds
- 150 Joules: < 4 seconds
- 200 Joules: < 6 seconds
- 270 Joules: < 7 seconds

  • Pads size

- Adult= area: 82 cm2 / 12.71in2
- Child= area: 30 cm2 / 4.65 in2

  • Maximum output voltage:

- 2000 V

  • Maximum output current:

- 60A (25 ohms)